瓶裝生物指示劑(ATCC 7953)

(訂購編號 GS/6)  瓶裝生物指示劑Geobacillus Stearothermophilus 106    ATCC 7953    50/

ATCC 7953

Geobacillus Stearothermophilus 106  

專供(Autoclave)蒸汽滅菌週期使用

試驗(Test): 有顏色改變(由紫色變黃色)或顏色有點變黃則指示出一個失敗的滅菌週期。 試劑瓶保持原來的紫色沒有混濁物或顏色改變的話,則指示出完成了適當的滅菌週期。

GS/6 瓶裝生物指示劑

 

產品消息

Biological Indicators

生物滅菌確效指示劑菌種名稱更改消息

自91年1月起ATCC

Bacillus stearothermophilus 更改為 Geobacillus stearothermophilus

Bacillus subtilis var niger 更改為 Bacillus atrophaeus.

謝謝大家的惠顧!  謝謝!

 

 

 

產品附中文使用說明書

Product include Chinese operation manual

 

 

BIOLOGICAL INDICATORS

         Self-Contained            

CERTIFICATE OF ANALYSIS     

Reorder No. GS/6                                               
Geobacillus stearothermophilus (Bacillus stearothermophilus)   ATCC 7953

For:       Steam Sterilization

Culture: GSTest Media, 55-60°C. The supplied bacteriological medium will

meet requirements for growth promoting ability.

Storage: Controlled room temperature. Store away from sterilants. Do not

desiccate.

Disposal: Incinerate or autoclave at 121°C for not less than 30 minutes.

Purity:      No evidence of contaminants using standard plate count techniques.

Lot No.:    S-276P

Manufacture Date: 2002 SEPTEMBER 23

Expiration: 24 months from Manufacture Date.  

Heat Shocked Population:     1.7x106              Spores/Unit

Assayed Resistance:

Temperature D-value(1) Survival Kill
121°C 1.9 8.0(2) 19.5(2)  minutes
132°C 0.3 1.0(3) 2.0  (3)  minutes
134°C 0.2 0.8(3) 2.0  (3)  minutes
135°C 0.2 0.6(3) 1.8  (3)  minutes
Z-value  11.9°C

D-value reproducible only when exposed In an AAMI BIER vessel and cultured under the exact conditions used to obtain results reported here. MPN method used.

Units are manufactured in compliance with FUZ Biotech's quality standards, USP, EN 866 and ISO 11138 guidelines and all appropriate subsections.

(1) D-value calculated using the Spearman-Karber method.

(2) Survival/Kill values are calculated according to USP and ISO 11138.

(3) Empirically derived data.

 

Monitoring Frequency: 

For optimum control of hospital sterilized goods, we recommend that an GSTest biological indicator be used to monitor every load of steam sterilized supplies containing implantable products or at least once a day for each steam autoclave.

Instructions For Use:

CAUTION: After sterilization, the contents of the GSTest biological indicator are hot and under pressure. Always allow to cool at least 10 minutes. Failure to cool at least 10 minutes may cause the glass ampoule to burst and may result in Injury from hot liquid.

A. Exposure:

  1. Remove an appropriate number of GSTest units from the box.

  2. Identify the indicators by labeling pertinent process information.

  3. Place an GSTest biological Indicator in a suitable test pack which is representative of the load (e.g., a linen pack for the load of linen, a tray of instruments for metal goods may be used).  

  4. Place this test pack in the most challenging area of the sterilizer, generally on the bottom shelf near the door over the drain. NOTE: If a test pack Is not being used, the GSTest unit should be oriented in a horizontal position during load processing.

  5. Process the load as usual.

  6. Remove from the sterilizer and allow the pack to cool for a sufficient time to cool the indicator, at least 10 minutes.

  7. Retrieve the GSTest biological indicator from the test load.

  8. The chemical indicator on the label changes from blue to black when exposed to steam. This distinguishes exposed from unexposed units.   NOTE: a black color does not indicate acceptable sterilization.

B. Incubation:

Any microbiological incubator that Is adjusted for 55 to 60°C will satisfy the incubation conditions for the GSTest. To activate the media, place the indicator in an upright position in a plastic crusher. Gently squeeze the crusher to break the glass ampoule. Place the activated indicator In the incubator rack, and incubate Immediately.

C. Interpretation

  1. Examine the indicator at regular intervals for any color change (i.e. 18 hours, 24 hours, 48 hours). The appearance of a yellow color indicates bacterial growth. No color change indicates adequate sterilization.

  2. Act on a positive test (a color change of yellow) as soon as the color change is noted. Notify appropriate hospital personnel (i.e. Infection Control). Always retest the sterilizer with several GSTest biological Indicators throughout the test load.

  3. The recommended incubation time Is 48 hours.

  4. Record the results.

  5. Dispose of all used GSTest biological Indicators in accordance with your institution's policy. Incinerate or autoclave any positive cultures at 250F (121 °C) for not less than 30 minutes.

Use of Controls

  1. As a positive growth control, place an activated,  unsterililzed GSTest biological indicator in each incubator on a daily basis.

  2. Examine the positive Indicator at regular intervals such as 18 and 24 hours. The yellow color is evidence of bacterial growth. Record the results. Remove all positive indicators as the yellow color is noticed, and dispose of as mentioned above.

  3. If the positive control does not grow, do not use the units from this box. Contact FUZ Biotech.

Storage:

  1. Store GSTest biological indicators at controlled room temperature conditions (15°C to 30°C) and 20 to 70 percent relative humidity.

  2. Do not store these Indicators near sterilants or other chemicals.

  3. GSTest biological indicators have a 24 month shelf life which Is clearly designated on each box. Rotate your stock accordingly.